Line 29: Line 29:
  
 Under 2D dynamic conditions, the following holds:\\ Under 2D dynamic conditions, the following holds:\\
-{{:CalTesterPlus_Eq4_v2.jpg}}\\+{{:CalTesterPlus_Eq4_v2.jpg}} \\
 \\ \\
 Rewriting the left-hand side of Eq. (4)\\ Rewriting the left-hand side of Eq. (4)\\
-{{:CalTesterPlus_Eq5_v2.jpg}}\\+{{:20140107212337caltesterplus_eq5_v2.jpg}} \\
 \\ \\
 Rearranging Eq. (5), the magnitude of the angular displacement ( {{:Beta.jpg}} ) between vectors r and A due exclusively to the inertial terms can be isolated:\\ Rearranging Eq. (5), the magnitude of the angular displacement ( {{:Beta.jpg}} ) between vectors r and A due exclusively to the inertial terms can be isolated:\\
 \\ \\
-{{:CalTesterPlus_Eq6_v2.jpg}}\\+{{:CalTesterPlus_Eq6_v2.jpg}} \\
 \\ \\
 where r is the length of the testing device rod, Icg the moment of inertia of the test device rod about the center of mass location, m the mass of the testing device rod and {{:ThetaDoubleDot.jpg}} is the angular acceleration of the testing device rod relative to an inertial reference frame. where r is the length of the testing device rod, Icg the moment of inertia of the test device rod about the center of mass location, m the mass of the testing device rod and {{:ThetaDoubleDot.jpg}} is the angular acceleration of the testing device rod relative to an inertial reference frame.