FP CALMATRIX
Jump to navigation
Jump to search
The CalMatrix premultiplies the analog channels converting the analog signals from volts to Newtons and/or Newton Meters (depending on the type of platform).
C3D Channels
- [Original Analog]= [Fx12, Fx34, Fy14, Fy23, Fz1, Fz2, Fz3, Fz4]
- as specified by the ANALOG channel numbers
Compute Baselines
- Average the Analog signals for the specified C3D ZERO frames
Subtract Baselines
- [Tared Analog]= [Original Analog]-[Baseline]
Processed Analogs
- If Processed Analog signals are used
- [Tared Analog]= [Processed Analog]-[Baseline]
Pre-Multiply by the CalMatrix
- [Analog]= [Fx12, Fx34, Fy14, Fy23, Fz1, Fz2, Fz3, Fz4]= [CalMatrix][Tared Analog]