The CalMatrix premultiplies the analog channels converting the analog signals from volts to Newtons and/or Newton Meters (depending on the type of platform).

C3D Channels

[Original Analog]= [Fx12, Fx34, Fy14, Fy23, Fz1, Fz2, Fz3, Fz4]
as specified by the ANALOG channel numbers

Compute Baselines

Average the Analog signals for the specified C3D ZERO frames

Subtract Baselines

[Tared Analog]= [Original Analog]-[Baseline]

Processed Analogs

If Processed Analog signals are used
[Tared Analog]= [Processed Analog]-[Baseline]

Pre-Multiply by the CalMatrix

[Analog]= [Fx12, Fx34, Fy14, Fy23, Fz1, Fz2, Fz3, Fz4]= [CalMatrix][Tared Analog]